风力发电用电缆扭转试验机温度校准方法

    Characterization and Calibration of Dual-Zone Temperature for Cable Torsion Testing Machines Used in Wind Power Generation

    • 摘要: 为研究风力发电用电缆扭转试验机的温度校准方法,提出一种仅测量设备底部区域与顶部区域温度进而评估设备温度是否符合要求的方法。分别在温度为60、−15、−40、−60 ℃的条件下,对型号为FLN300的风力发电用电缆扭转试验机进行温度校准。结果表明,不同温度下,温度最大偏差都出现在试验设备的顶部区域;设备温度偏差由顶部向底部呈现逐渐减小的趋势。在不同温度下,当设备顶部区域温度符合标准要求时,设备内部的温度偏差均满足标准要求。文中的温度校准验证过程为该类设备温度校准提供了参考,简化了设备校准过程。

       

      Abstract: To study temperature calibration methods for torsion testing machines used for wind power cables, a method was proposed that assessed whether equipment temperature met requirements by measuring only the temperatures in the bottom and top regions of the equipment. Temperature calibration was performed on a FLN300 wind power cable torsion testing machine at temperatures of 60 °C, −15 °C, −40 °C, and −60 °C. The results showed that at different temperatures, the maximum temperature deviation occurred in the top region of the testing equipment, and the temperature deviation decreased gradually from the top to the bottom. At different temperatures, when the temperature in the top region of the equipment met the standard requirements, the internal temperature deviation of the equipment also satisfied the standard requirements. The temperature calibration verification process described in this paper provides a reference for the temperature calibration of such equipment and simplifies the calibration procedure.

       